Patient-protective Radiation Protection Shields Used During Computed Tomography (CT) Examinations Reduce the Scattered Dose in CT Room

https://kwmed.repo.nii.ac.jp/records/2000133
https://kwmed.repo.nii.ac.jp/records/2000133
5dd78be6-ce14-4476-8a99-85e88581120e
名前 / ファイル ライセンス アクション
KMJ-E202450025.pdf PDF (1.3 MB)
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2025-06-19
タイトル
タイトル Patient-protective Radiation Protection Shields Used During Computed Tomography (CT) Examinations Reduce the Scattered Dose in CT Room
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 ARAO, Shinichi

× ARAO, Shinichi

en ARAO, Shinichi

Search repository
MASUDA, Takanori

× MASUDA, Takanori

en MASUDA, Takanori

Search repository
OKU, Takayuki

× OKU, Takayuki

en OKU, Takayuki

Search repository
ONO, Atsushi

× ONO, Atsushi

en ONO, Atsushi

Search repository
OKURA, Yasuhiko

× OKURA, Yasuhiko

en OKURA, Yasuhiko

Search repository
著者所属(英)
言語 en
値 Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare
著者所属(英)
言語 en
値 Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare
著者所属(英)
言語 en
値 Department of Medical Assistance Radiology, Tsuchiya General Hospital
著者所属(英)
言語 en
値 Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare
著者所属(英)
言語 en
値 Department of Clinical Radiology, Faculty of Health Sciences, Hiroshima International University
キーワード
言語 en
主題Scheme Other
主題 Radiation protection shields
キーワード
言語 en
主題Scheme Other
主題 Triple-phase liver CT
キーワード
言語 en
主題Scheme Other
主題 Healthcare workers
キーワード
言語 en
主題Scheme Other
主題 Scattered dose
抄録(英)
言語 en
値 Objective: This study aimed to conduct a comparative analysis of the computed tomography (CT) scatter radiation dose healthcare workers are exposed to with and without the radiation protection shields (RPS).
Methods: A helical scan of a routine triple-p se liver CT protocol was performed using a tissue-equivalent phantom via a 64-detector-row CT scanner. Scattered doses were measured and compared with and without RPS wrapped around the neck to chest outside the imaging area using a calibrated pocket electron dosimeter. We measured at 50 cm, 100 cm and 150 cm above the floor at a total of 57 points in front, to the side and behind the gantry, at 50 cm intervals from the centre of the gantry. The heights of 50 cm, 100 cm and 150 cm from the floor assume the estimated positions of the gonads, mammary glands and lens, respectively.
Results: Using RPS outside the imaging area reduced the scattered dose by about 10% at all measurement positions and heights (p < 0.05).
Conclusion: The use of RPS on patients during CT examinations reduces the scattered dose in the CT room by approximately 10%.
書誌情報 en : Kawasaki medical journal

巻 50, p. 25-31, 発行日 2024
出版者
出版者 Kawasaki Medical Society
言語 en
ISSN
収録物識別子タイプ PISSN
収録物識別子 03850234
ISSN
収録物識別子タイプ EISSN
収録物識別子 24343404
書誌レコードID
収録物識別子タイプ NCID
収録物識別子 AA12029005
DOI
関連タイプ isIdenticalTo
識別子タイプ DOI
関連識別子 https://doi.org/10.11482/KMJ-E202450025.pdf
記事種別(英)
内容記述 Regular Article
言語 en
関連サイト
識別子タイプ URI
関連識別子 https://igakkai.kms-igakkai.com/wp/wp-content/uploads/2024en/KMJ-E202450025.pdf
著者版フラグ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
戻る
0
views
See details
Views

Versions

Ver.1 2025-06-18 07:34:29.900393
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3