タイトル |
|
|
タイトル |
Development of a screening system for central visual field using the eye-tracking device\n |
|
言語 |
en |
言語 |
|
|
言語 |
eng |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
著者 |
FUJIWARA, Atsushi
MIKI, Atsushi
ARAKI, Syunsuke
OKA, Mayumi
YONEDA, Tsuyoshi
YAOEDA, Kiyoshi
YAMASHITA, Tsutomu
OUCHI, Tatsuhiro
MARUHISA, Yuriko
FUJITA, Mika \n
|
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital・Department of Ophthalmology, Kawasaki Medical School |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Ophthalmology, Kawasaki Medical School Hospital |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Yaoeda Eye Clinic |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Doctoral Program in Sensory Science, Graduate School of Health Science and Technology, Kawasaki University of Medical Welfare |
著者所属(英) |
|
|
言語 |
en |
|
値 |
Department of Orthoptics, Faculty of Rehabilitation, Kawasaki University of Medical Welfare・Department of Ophthalmology, Kawasaki Medical School Hospital |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
Visual field |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
Gaze movement |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
Eye-tracking device |
抄録(英) |
|
|
言語 |
en |
|
値 |
Background: Visual field test with gaze movements do not require a subjective response because they are based on reflexive movements. In this study, we developed a visual field test system with gaze movements to perform a central visual field screening, and then examined the reproducibility of the measurements in healthy adult volunteers. Methods: We examined 30 right eyes of 30 healthy volunteers (mean age, 22.7 ± 5.2 years) with a best-corrected visual acuity of at least 20/20. Gazefinder, an eye-tracking device, was used to measure gaze movements. Subjects with refractive correction were asked to follow a white target presented on a monitor. If a subject can accurately perform eye tracking with respect to the visual target, visual field with gaze movements measurements are theoretically possible in eight directions (horizontal/vertical to 15.3° and oblique to 21.5° ). After a total of three measurements, the data were quantified using analysis software (CreateChart). Finally, the intraclass correlation coefficients of the measurement values were obtained. Results: The difference between theoretical and actual measurement values, which is thought to reflect gaze accuracy, were –0.1° ± 0.9° for upper, –0.6° ± 1.0° for upper right, –0.2° ± 1.0° for right, –0.8° ± 0.9° for lower right, –0.5° ± 0.7° for lower, –0.5° ± 0.9° for lower left, –0.6 ° ± 0.5 ° for left, and –0.6 ° ± 0.5 ° for upper left. No significant differences were found among the eight directions, and gaze accuracy was high, at within 1°. The intraclass correlation coefficients were 0.6 or higher in each direction (P < 0.01), indicating high repeatability. Conclusions: In the traditional method for measuring visual field with gaze movements, the fixation point of view needs to be reset for each gaze movement. On the other hand, the system developed in this study has the advantage of not requiring eye movements to return to the fixation point. The present findings indicate that our newly developed system is a useful device when standard perimetry is difficult to measure. |
書誌情報 |
en : Kawasaki medical journal
巻 48,
p. 191-199,
発行日 2022
|
出版者 |
|
|
出版者 |
Kawasaki Medical Society |
ISSN |
|
|
収録物識別子タイプ |
EISSN |
|
収録物識別子 |
24343404 |
書誌レコードID |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AA12029005 |
DOI |
|
|
関連タイプ |
isIdenticalTo |
|
|
識別子タイプ |
DOI |
|
|
関連識別子 |
https://doi.org/10.11482/KMJ-E202248191 |
記事種別(英) |
|
|
内容記述 |
Regular Article |
|
言語 |
en |
関連サイト |
|
|
|
識別子タイプ |
URI |
|
|
関連識別子 |
https://igakkai.kms-igakkai.com/wp/wp-content/uploads/2022en/KMJ-E202248191.pdf |
著者版フラグ |
|
|
出版タイプ |
VoR |
|
出版タイプResource |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |