A Case of False Aneurysm of a Dacron Graft in Aortitis Syndrome

https://kwmed.repo.nii.ac.jp/records/521
https://kwmed.repo.nii.ac.jp/records/521
76e4d936-f18a-44ec-8bad-c12c51dde41f
名前 / ファイル ライセンス アクション
KJ00000010569.pdf KJ00000010569.pdf (1.3 MB)
Item type [ELS]学術雑誌論文 / Journal Article(1)
公開日 2017-01-23
タイトル
タイトル A Case of False Aneurysm of a Dacron Graft in Aortitis Syndrome
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 MORITA, Ichiro

× MORITA, Ichiro

en MORITA, Ichiro

Search repository
MASAKI, Hisao

× MASAKI, Hisao

en MASAKI, Hisao

Search repository
ISHIDA, Atsuhisa

× ISHIDA, Atsuhisa

en ISHIDA, Atsuhisa

Search repository
TABUCHI, Atsushi

× TABUCHI, Atsushi

en TABUCHI, Atsushi

Search repository
KUBO, Hiroshi

× KUBO, Hiroshi

en KUBO, Hiroshi

Search repository
TANEMOTO, Kazuo

× TANEMOTO, Kazuo

en TANEMOTO, Kazuo

Search repository
著者所属(英)
en
Department of Thoracic and Cardiovascular Surgery, Kawasaki Medical School
著者所属(英)
en
Department of Thoracic and Cardiovascular Surgery, Kawasaki Medical School
著者所属(英)
en
Department of Thoracic and Cardiovascular Surgery, Kawasaki Medical School
著者所属(英)
en
Department of Thoracic and Cardiovascular Surgery, Kawasaki Medical School
著者所属(英)
en
Department of Thoracic and Cardiovascular Surgery, Kawasaki Medical School
著者所属(英)
en
Department of Thoracic and Cardiovascular Surgery, Kawasaki Medical School
キーワード
言語 en
主題Scheme Other
主題 Dacron Graft
キーワード
言語 en
主題Scheme Other
主題 Nonanastomotic aneurysm
抄録(英)
en
Among the complications after grafting of a prosthesis, false aneurysm due to a torn prosthesis is extremely rare. A 56-year-old female was operated on for complete occlusion of the left external iliac artery. A bypass was performed from the left common iliac artery to the left femoral artery in 1971, using a double-velour Dacron graft (Wesolowski) 8 mm in diameter. In August 1996, the patient noticed a pulsatile mass in her left groin. Angiography revealed a false aneurysm at the site of the graft. On operation, the false aneurysm was found to have formed on the mid-distal-portion of the graft. After resection of the aneurysm, reconstruction was successfully achieved with an 8 mm Gelsoft double velour knitted Dacron graft (It common iliac-femoral bypass). The postoperative course was uneventful. Postoperative angiography showed a well-patented graft. Fractured fibers were noted in the aneurysmal part of the resected specimen. Therefore, the false aneurysm was caused by rupture of the implanted graft.
書誌情報 en : Kawasaki medical journal

巻 28, 号 2・3・4, p. 87-92, 発行日 2002
URL
識別子 http://igakkai.kms-igakkai.com/wp/wp-content/uploads/2002en/28(2.3.4)87-92.2002.pdf
識別子タイプ URI
DOI
識別子タイプ DOI
関連識別子 https://doi.org/10.11482/KMJ28(2.3.4)87-92.2002.pdf
ISSN
収録物識別子タイプ PISSN
収録物識別子 0385-0234
ISSN
収録物識別子タイプ EISSN
収録物識別子 2434-3404
雑誌書誌ID
収録物識別子タイプ NCID
収録物識別子 AA00710169
雑誌書誌ID
収録物識別子タイプ NCID
収録物識別子 AA12029005
著者版フラグ
出版タイプ VoR
戻る
0
views
See details
Views

Versions

Ver.1 2023-06-19 11:05:21.010858
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3