Experimental validation of blood flow derived from pulse oximeter wave signals in beagles

https://kwmed.repo.nii.ac.jp/records/576
https://kwmed.repo.nii.ac.jp/records/576
cdc2d11b-4f7e-4340-83a5-6d49342426a3
名前 / ファイル ライセンス アクション
KJ00009888239.pdf KJ00009888239.pdf (781.8 kB)
Item type [ELS]学術雑誌論文 / Journal Article(1)
公開日 2017-01-23
タイトル
タイトル Experimental validation of blood flow derived from pulse oximeter wave signals in beagles
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Yamamoto, Masako

× Yamamoto, Masako

en Yamamoto, Masako

Search repository
Fujita, Yoshihisa

× Fujita, Yoshihisa

en Fujita, Yoshihisa

Search repository
Sakuta, Yuka

× Sakuta, Yuka

en Sakuta, Yuka

Search repository
Tsuge, Masatsugu

× Tsuge, Masatsugu

en Tsuge, Masatsugu

Search repository
著者所属(英)
en
Departments of Anesthesiology & ICM Kawasaki Medical School
著者所属(英)
en
Departments of Anesthesiology & ICM Kawasaki Medical School
著者所属(英)
en
Departments of Anesthesiology & ICM Kawasaki Medical School
著者所属(英)
en
Departments of Anesthesiology & ICM Kawasaki Medical School
キーワード
言語 en
主題Scheme Other
主題 Pulse oximetry
キーワード
言語 en
主題Scheme Other
主題 Blood flow
キーワード
言語 en
主題Scheme Other
主題 Monitor
記事種別(英)
内容記述タイプ Other
内容記述 Original Article
抄録(英)
en
Purpose Pulse oximeter wave reflects blood volume changes in tissue, suggesting the possibility of monitoring changes in tissue blood flow. Thus, our aim was to examine the correlation between tissue blood flow derived from pulse oximeter wave signals (Q_pulse) at a toe and the arterial flow measured by a Doppler probe at the femoral artery (Q_Doppler).Methods Six beagles under general anesthesia were studied. A 24-G catheter was placed in the proximal femoral artery for drug infusion and an ultrasonic transit-time flow probe applied to the artery to measure Q_Doppler. The pulse oximeter signals from the right toe were processed with in-house-developed software to obtain Q_pulse. Three saline solutions containing respectively the vasodilators isosorbide dinitrate (20μg/mL), adenosine (20μg/mL), and nicardipine (10μg/mL) were infused at increasing rates of 0, 2.5, 5.0, 10, or 20 mL/h for 8 minutes into the femoral artery with a syringe pump. Results Both Q_Doppler and Q_pulse increased fourfold with increasing rates of infusion of the three vasodilators. Plotting of Q_Doppler and Q_pulse across the three vasodilators in each animal revealed linear correlations (R^2 = 0.17-0.76). Overall regression analysis showed a less strong but still statistically significant linear relation (y=3.68x + 18.5, R^2 = 0.25, P < 0.01). Conclusions We found a linear correlation between Q_Doppler and Q_pulse in a wide range of femoral arterial blood flow measures induced by different vasodilators in each animal. Arterial flow wave derived from pulse oximetry was quantitatively validated.
書誌情報 en : Kawasaki medical journal

巻 41, 号 1, p. 1-5, 発行日 2015
URL
識別子 http://igakkai.kms-igakkai.com/wp/wp-content/uploads/2015en/KMJ-E41(1)1.pdf
識別子タイプ URI
DOI
識別子タイプ DOI
関連識別子 https://doi.org/10.11482/KMJ-E41(1)1
ISSN
収録物識別子タイプ PISSN
収録物識別子 0385-0234
ISSN
収録物識別子タイプ EISSN
収録物識別子 2434-3404
雑誌書誌ID
収録物識別子タイプ NCID
収録物識別子 AA12685295
雑誌書誌ID
収録物識別子タイプ NCID
収録物識別子 AA12029005
著者版フラグ
出版タイプ VoR
戻る
0
views
See details
Views

Versions

Ver.1 2023-06-19 11:07:31.934764
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3